Description
Specification :-
Anesthesia Workstation, with its ergonomic design, integrates a concise and smooth structure and high mobility.
The microcomputer-controlled Ventilator adopts exported sensors and other components. Its scientific and proper design as well as rigorous production process are what demonstrate the inner quality of product. Mechanical ventilation is made secure and visible by the folded respiratory bellows, and the security of air supply is guaranteed by pneumatic transmission and measurement system. With multiple modern technologies applied, the Anesthesia Workstation is both outstanding in performance and versatile in functions.
The Ventilator adopts microcomputer control with high-definition 5.7” LCD display, integrating the functions of respiratory control and displaying the monitoring parameters. It is versatile in function and excellent in performance.
The high Precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose two kinds of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
High-definition four-tube flowmeter, designed with stoppers for oxygen and nitrous oxide to control the output oxygen concentration.
Reliable pneumatic transmission system.
Main Specification of Anesthesia Workstation :
Main body : High-strength engineering plastic rack, light, beautiful and corrosion resistant
Scope of application : Adult and Child
Gas source : O2: 0.28~0.6MPa N2O: 0.28~0.6MPa
Flow meter : O2: 0.05~1.0L/min 1.1~10L/min
N2O: 0.05~1.0L/min 1.1~10L/min
O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
Flow rate of rapid oxygen supply : 25~75L/min
Low oxygen pressure alarm : There will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer : It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%.
Among Halothane,Enflurane,Isoflurane and Sevoflurane, two can be chosen for application as required by the customer.
Respiratory circuit working mode : All-close, semi-close, semi-open APL ≤12.5kpa
Respiratory Bellows : Bellows for adults, bellows for children tidal volume range:0~1,500 ml
Main Specification of Ventilator
Display mode : High-definition 5.7” LCD screen display
Ventilation mode : IPPV, SIPPV, SIMV, MANUAL
Ventilation function : PEEP,SIGH, IRV,
Ventilation parameters : Tidal volume 50~1500 ml Rate 2~99 bpm
SIMV rate 2~20 bpm I:E 2:1~1:8
Inspiratory trigger pressure -10~20cmH2O PEEP 2~20cm H2O Pressure Range5~60cmH2O
SIGH1.5 times the inspiratory time (60-120 adjustable)
Parameters for ventilation monitoring tidal volume, minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure – time waveform, flow rate – time
waveform, PEEP, inspiratory trigger pressure
Security Alarm System
Oxygen Concentration Alarm : Upper limit setting range (21%~99%)
Low limit setting range (18%~80%)
Airway Pressure Alarm : Upper limit setting range 5~60 cm H2O
Low limit setting range 0~50 cmH2O
Minute Ventilation Volume Alarm : Upper limit setting range 1~20 L/min
Low limit setting range 0~19L/min
Tidal Volume : Upper Limit Setting Range 100~1500ml
Low Limit Setting Range : 0~1400ml
Sustained high-pressure alarm : It will give alarm when stress have consistently been higher than 2.5cm H2O
(Suffocation alarm, Power alarm, Intubation off alarm, battery voltage alarm)
Power : AC 220V 50Hz
(Mechanical Arms End-tidal CO2 monitoring)
We are Exporters and Suppliers of Anaesthesia Workstation Gold at wholesale competitive prices from India.
Model No :
APAW 111 Anaesthesia Workstation Gold
Reviews